Operating & Maintenance ManualAlert-3 LCD Alarm v2.0 www.amico.com1ST FLOOR ROOM 122ND FLOORROOM 253RD FLOOR ROOM 37MAINTEN EXT 3024TH FLOORROOM 4
10 Amico PipelineSTEP 7: SENSOR MODULEB: LOCAL (inside the back box)i. The sensor module is provided with a 6”-8” [0.1 m - 0.2 m] twisted pair of
www.amico.com 11STEP 9: LCD DISPLAY SETUPi. Press the SETUP button (B1) and press the SELECT button (B4)ii. Volume control: 90, 80, 70, 60
12 Amico PipelineProgramming Gas Locations1. Use the Notepad program to enter information for gas location. Each line can hold up to a maximum of
www.amico.com 13Model Numbers “L” Denes the Language:U = English (NFPA)E = English (CSA/ISO)F = French (CSA/ISO)S = Spanish (NFPA/ISO)“Gas”
14 Amico PipelineSpare Part NumbersModel Number DescriptionA3P-SENS-E-AIRSensor Module ISO-AIR Eng. Alert-3A3P-SENS-E-CO2Sensor Module ISO-CO2 En
www.amico.com 15Spare Part NumbersModel Number DescriptionS-DIS-DEMC-AIR DISS Demand Check Valve 1/4” MNPT - AIRS-DIS-DEMC-CO2 DISS Demand C
16 Amico PipelineDimensionsLCD ALARM1/2”-14 NPSM [13] (Approximate, can vary)Wrench Flats1[25]1/2” ConduitConnection#22 Gauge twisted pairshielded
www.amico.com 17TroubleshootingSymptom Cause Corrective ActionAn error or “LOW ALARM” LCD screen The Microprocessor detected a fault and has
18 Amico PipelineTroubleshootingFACTORY DEFAULT SETTING GASMid Pressure Hi = 60 psi Low = 40 psiVacuum Hi = 32 inHg L
www.amico.com 19Appendix AWIRING DIAGRAM: LCD MOTHERBOARDB4 B3 B2 B1BuzzerLCD CableConnectorSensor TerminalMute Button(located on reverse sid
2 Amico PipelineContentsUser Responsibility 3Introduction 3 - 4Features Description of the Alarm 5 - 6Shipment Details 5The Alarm Back Box
20 Amico PipelineAppendix BWIRING DIAGRAM: AUTO-SWITCH POWER SUPPLYAC Supply115 to 220 VACDC Power Cable:Connect toAnnunciator ModuleGNLL - LiveN
www.amico.com 21Appendix CWIRING DIAGRAM: LCD DISPLAY MODULE - ALARM BUZZERNOTE:Relays on the annunciator are failsafe for Version 3.1 or new
22 Amico PipelineAppendix DWIRING DIAGRAM: LCD DISPLAY MODULE - LOCAL SENSORB4 B3 B2 B1SETUPCHANGE UPCHANGE DOWNSELECT/MUTEAC Supply115 to 220 VAC
www.amico.com 23WIRING DIAGRAM: LCD DISPLAY MODULE - REMOTE SENSORNote:For multiple sensors, amulti-conductor twistedpair cable can be used.O
24 Amico PipelineAppendix FWIRING DIAGRAM: LCD DISPLAY MODULE - MASTER MODULE#22 Gauge twistedpair shielded cable6"-8” [0.1 m-0.2 m] supplied
www.amico.com 25TECHNICAL SPECIFICATIONSSupply Voltage: 115 - 220 VAC, 50 - 60 HzCurrent Draw: 1 Amp. Max.Fuse (1/4 * 1-1/4): Fast Blow
26 Amico Pipeline
www.amico.com 27
Amico Pipeline | www.amico.com 85 Fulton Way, Richmond HillOntario, L4B 2N4, CanadaToll Free Tel: 1.877.462.6426Toll Free Fax: 1.866.440.4986Tel: 905.
www.amico.com 3ContentsTroubleshooting 17 - 19Appendix A 20Wiring Diagram: LCD MotherboardAppendix B 21Wiring Diagram: Auto-Switch Power S
4 Amico PipelineUser ResponsibilityIntroductionThe information contained in this Installation, Operation and Maintenance Manual pertains only to t
www.amico.com 5FeaturesDescription of the Alarm• Microprocessor based digital LCD and individual microprocessor on each sensor module• Gas-
6 Amico PipelineDescription of ModulesThe Alert-3 LCD Alarm is a high technology microprocessor based module: COMMON TO ALL ALARMSSYSTEM POWER SUP
www.amico.com 7Installation GuideSTEP 1: THE ALARM BOXInstall the back-box to the studs of the wall at the desired height. Ensure that the bo
8 Amico PipelineAmico Microprocessor Based AlarmPage: 12CAUTION: The microprocessor circuitry on the ALERT-2 alarm con-tains sophisticated integra
www.amico.com 9STEP 6: SYSTEM POWER SUPPLY TURN OFF POWER SWITCH before changing any modules and/or disconnecting any cables. Failure to do s